PROCESS PHOTOMETRY

 

No.  14003-4R/N

 

PHOSPHATE Total ( TP ) : UV digestable molibdenum blue method

Range:  0.2 –  10  mg/l ( P  tot. )

 

PRINCIPLE

 

The automated procedure for the determination of total Phosphate involve the step of UV digestable with an excess of potassium persulfate . The organic phosphate compound present in the sample, are converted into o-phosphate. All phosphate present in o-phosphate form, is then  determined with molybdenum blue method

The color intensity for molybdenum blue method is measured at 660 nm